Medical devices

EN 1060

Non-invasive sphygmomanometers (blood pressure monitors)

The standard for blood pressure measuring devices, accuracy, safety and clinical validation.

The EN 1060 series sets the requirements for non-invasive sphygmomanometers — devices measuring arterial pressure through a cuff, without entering a blood vessel.

Parts of the series

1060-1
General requirements
1060-2
Supplementary requirements for mechanical devices
1060-3
Supplementary requirements for electro-mechanical systems
1060-4
Test procedures to determine overall system accuracy

The accuracy requirement

The maximum permissible error is ±3 mmHg (or ±2%, whichever is greater). That limit matters: a systematic 5 mmHg error would materially skew hypertension diagnosis at population level.

The occupational health context

In an industrial workplace, a blood pressure monitor is often part of the medical room and the periodic health surveillance programme. Its calibration must be checked periodically — a mechanical gauge drifts over time, and a cuff that does not match the arm circumference introduces a systematic error.